ERR 2 — clock failure
JOWA SEAGUARD, 15 ppm bilge alarm (ows monitor)
Possible cause
Hardware failure
What to do
Manufacturer's manual, p. 22Switch off, remove the battery from the control unit for 24 h, switch on, set GMT time and date, refit the battery; otherwise replace the CPU board
See the assembly on the 3D model
Other JOWA SEAGUARD faults
ERR 4 — memory errorMemory card loose or faulty (corroded contacts)ERR 8 — sensor not connectedBroken/loose cable or faulty sensorERR 10 — sensor cleaning requiredNo clean water, dirty sensor, failed auto-flush, air bubbles, faulty sensorERR 20 — sensor temperatureTemperature outside +1…+60 °C or faulty sensorERR 40 — log truncatedLog recording interrupted by power failureERR 80 — unknown errorSoftware failure or no ID in the Information menuPPM: O/R — out of rangeValue outside 0–30 ppm, sample too dirty, faulty sensor
From the manufacturer's manual “Manual JOWA SEAGUARD”. Check against the original on board before work.